Cas de réparation de dispositifs médicaux

Qualité Olympus CF-H190L Colonoscope pour réparation Usine

Olympus CF-H190L Colonoscope pour réparation

Olympus CF-H190L Colonoscope for repair ► Brief Description Fault: 1. Image ok; 2. Light bundle still ok; 3. The front end of the insert tube is broken, and the serpentine; the position of stress boot is leaking; 4. Biopsy tube wear; 5. Left Angle failure, bending section, wire and helix tube wear; ...
Qualité Sonde cardiaque Samsung Medison P2-4AH pour réparation Usine

Sonde cardiaque Samsung Medison P2-4AH pour réparation

Samsung Medison P2-4AH Cardiac probe for repair ► Brief Description Fault:Lens problem, crystal damage. Solution:Replace probe lens, replace crystal. ►Problem & Solutoin Brand Samsung Model P2-4AH Fault Lens problem, crystal damage Solution Replace probe lens, replace crystal ► Common issues with P2...
Qualité DUODÉNOSCOPE OLYMPUS TJF-Q180V POUR RÉPARATION Usine

DUODÉNOSCOPE OLYMPUS TJF-Q180V POUR RÉPARATION

OLYMPUS TJF-Q180V DUODENOSCOPE FOR REPAIR ► Brief Description Fault: 1. Image ok; 2. Light bundlestill ok; 3. Insert tubestill ok; 4. Biopsy tubewear; 5. Bendingsection, wires and helix tube wear; 6. Knob leakage. Solution: Suggest toreplace biopsy tube, bending section, wires, helix tube, knob and ...
Qualité Olympus CF-240L Colonoscope pour réparation Usine

Olympus CF-240L Colonoscope pour réparation

Olympus CF-240L Colonoscope for repair ► Brief Description Fault: Leakage, glass of light bundle blur, rubber leak, angle not enough,image fuzzle,objective lens scratch. Solution: Replace rubber, objective lens, C cover. Adjust the angle. ►Problem & Solutoin Brand Olympus Model CF-240L Fault Leakage...
Qualité Endoscope Karl Storz 27005BA pour réparation Usine

Endoscope Karl Storz 27005BA pour réparation

Karl Storz 27005BA endoscope for repair ► Brief Description Fault: 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens ...
Qualité Sonde à réseau phasé Toshiba PST-25BT pour réparation Usine

Sonde à réseau phasé Toshiba PST-25BT pour réparation

Toshiba PST-25BT Phased Array Probe For Repair ► Brief Description Fault: Housing break, image weak, crystal aging. Solution: Replace crystal and housing. ►Problem & Solutoin Brand Toshiba Model PST-25BT Fault Housing break, image weak, crystal aging Solution Replace crystal and housing ► Common ...
Qualité La tête de caméra STRYKER 1188-210-105 est en réparation. Usine

La tête de caméra STRYKER 1188-210-105 est en réparation.

STRYKER 1188-210-105 camera head for repair ► Brief Description Fault: Image with interrupt. Cable aging. Solution: Replace compatible new cable and connector. ►Problem & Solutoin Brand STRYKER Model 1188-210-105 Fault Image with interrupt. Cable aging Solution Replace compatible new cable and ...
Qualité La tête de la caméra STORZ H3-ZA à réparer Usine

La tête de la caméra STORZ H3-ZA à réparer

STORZ Camera Head H3-ZA For Repair ► Brief Description Fault:Image streaks frequently,cable loose contact and aged. Solution: Replace cable kit. ►Problem & Solutoin Brand STORZ Model H3-ZA Fault Image streaks frequently,cable loose contact and aged Solution Replace cable kit ► Common issues with H3...
Qualité GE RIC6-12-D sonde d'endocavité à réparer Usine

GE RIC6-12-D sonde d'endocavité à réparer

GE RIC6-12-D endocavity probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens,Clean cable. ►Problem & Solutoin Brand GE Model RIC6-12-D Fault Lens problem Solution Replace lens,Clean cable ► Common issues with RIC6-12-D Probe Dome Damage and Fluid Leaks Image Dropouts ...
Qualité La sonde à ultrasons Hitachi C41V1 est en réparation. Usine

La sonde à ultrasons Hitachi C41V1 est en réparation.

Hitachi C41V1 Ultrasuound probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens. ►Problem & Solutoin Brand Hitachi Model C41V1 Fault Lens problem Solution Replace lens ► Common issues with C41V1 Probe Image Dropouts or Streaks Cable and Connector Damage Acoustic Lens ...
Qualité Sonoscape 8-L743 sonde linéaire pour réparation Usine

Sonoscape 8-L743 sonde linéaire pour réparation

Sonoscape 8-L743 linear probe for repair ► Brief Description Fault: Crystal problem. Solution: Replace the Crystal. ►Problem & Solutoin Brand Sonoscape Model 8-L743 Fault Crystal problem Solution Replace the crystal ► Common issues with 8-L743 Probe Acoustic Lens Damage Piezoelectric Crystal Damage ...
Qualité Sonoscope C362 Probes abdominales à réparer Usine

Sonoscope C362 Probes abdominales à réparer

Sonoscape C362 Abdominal probe for repair ► Brief Description Fault: Cystal problem. Solution: Replace the Crystal. ►Problem & Solutoin Brand Sonoscape Model C362 Fault Cystal problem Solution Replace the crystal ► Common issues with C362 Probe Image Dropouts (Black Lines or Streaks) Damaged Lens ...
Qualité Sonde abdominale GE 4C-RS pour réparation Usine

Sonde abdominale GE 4C-RS pour réparation

GE 4C-RS abdominal probe for repair ► Brief Description Fault: Lens broken, cable broken, crystal bad. Solution: Replace probe head, repair cable. ►Problem & Solutoin Bran GE Model 4C-RS Fault Lens broken, cable broken, crystal bad Solution Replace probe head, repair cable ► Common issues with 4C-RS ...
Qualité Sonde échographique Careutech BK Medical 9048 pour réparation Usine

Sonde échographique Careutech BK Medical 9048 pour réparation

BK Medical 9048 Ultrasound Probe For Repair ► Brief Description Fault: The housing break. Solution: Replace probe housings. ►Problem & Solutoin Brand BK Medical Model 9048 Fault The housing break. Solution Replace probe housings. ► Common issues with 9048 Probes No Image / Poor Image Quality Probe ...
Qualité Pièces d' équipement médical Sonoscope L743 sonde linéaire pour réparation Usine

Pièces d' équipement médical Sonoscope L743 sonde linéaire pour réparation

Sonoscape L743 linear probe for repair ► Brief Description Fault: Probe head broken. Solution: Replace probe head.(including crystal and upper housing). ►Problem & Solutoin Brand Sonoscape Model L743 Fault Probe head broken Solution Replace probe head.(including crystal and upper housing) ► Common ...
Qualité Sonde échographique robotisée BK Medical 9096 pour réparation Usine

Sonde échographique robotisée BK Medical 9096 pour réparation

BK Medical 9096 Robotic Ultrasound Probe For Repair ► Brief Description Fault: All the outer skin of the cable is seriously damaged. Solution: Replace the outer skin of the cable. ►Problem & Solutoin Brand BK Medical Model 9096 Fault All the outer skin of the cable is seriously damaged Solution ...
Qualité Sonde Samsung Medison 3D4-9 SN P0N7  Pour réparation Usine

Sonde Samsung Medison 3D4-9 SN P0N7 Pour réparation

Samsung Medison 3D4-9 Probe SN P0N7 For Repair ► Brief Description Fault: Scanhead broken. Solution: Replace scanhead ► Problem & Solutoin Brand Samsung Model 3D4-9 Fault Scanhead broken Solution Replace scanhead ► Common i ssues with 3D4-9 Probes Cable damage Lens delamination Element failure (3D...
Qualité Cas de réparation de dispositifs médicaux de GE RIC5-9-D sonde à ultrasons pour réparation Usine

Cas de réparation de dispositifs médicaux de GE RIC5-9-D sonde à ultrasons pour réparation

GE RIC5-9-D Probe Fault: Dome loose, straight relief tiny broken, with error on machine. Solution: Replace dome and oiling. Repair cost : $***. Common Issues with the GE RIC5-9-D Ultrasound Probe: Image Artifacts/Dropouts: These can manifest as lines, shadowing, or a complete loss of image in ...
Qualité Fil d'angle et tube hélicoïdal usé CF-HQ190L 20% boîte de lumière cassée Usine

Fil d'angle et tube hélicoïdal usé CF-HQ190L 20% boîte de lumière cassée

Products: CF-HQ190L For Repair Problem & Solutoin: CF-HQ190L - Light bundle 20% Broken Knob Leaked The angle wire and helix tube Worn Biospy channel Worn Rubber Cracked Light guide tube Indented Solution Replace Biopsy Channel, Wire, Light Guide Tube, Knob, Rubber Applications: When it comes to ...
Qualité Cas de réparation de dispositifs médicaux de l' Olympus CF-H190L Colonoscope pour réparation Usine

Cas de réparation de dispositifs médicaux de l' Olympus CF-H190L Colonoscope pour réparation

Olympus CF-H190L Colonoscope for repair,SN:261****. Fault:Light bundle broken,biopsy tube worn,water air tube leakage,bending section deformation,wire helix tube worn. Solution:Replace the bending section,wire,helix tube,air tube and rubber.PS:Warranty for replace parts only. Repair time:7 working ...
Qualité Équipement médical Olympus réparé solution ultime pour les services médicaux Usine

Équipement médical Olympus réparé solution ultime pour les services médicaux

Product Description: Product Summary: The Medical Device Repair Cases product offered is a comprehensive solution for repairing and maintaining Olympus medical devices. With a warranty of 90 days, customers can have peace of mind knowing that their repaired devices are covered for an extended period...
Qualité OLYMPUS CF-HQ190L Casse de réparation de dispositifs médicaux Casse d'outil de réparation Usine

OLYMPUS CF-HQ190L Casse de réparation de dispositifs médicaux Casse d'outil de réparation

OLYMPUS CF-HQ190L Repair Problem: 1, Image display;2, Light bundle broken about 10%;3, Insertion tube is ok;4, Biospy channel worn and leaking;5, The upper angle broken, helix tube and bending section were worn;6, Light guide tube indented and the back cover is leaking;7, The EL-connector is not ...